JMSSJ On-line, Vol. 48 (2000) No. 2, pp. 155-159
Simultaneous Determination Method of Emedastine and Its Metabolites in Human Plasma Using High-Performance Liquid Chromatography-Tandem Mass Spectrometry
>> Full Text PDF
>> References
    Rie TAKAYAa), Mitsuhiro ISHIDAa)*, and Fumio SAKAMOTOa)

    *a) Research and Development Division, Nippon Organon K. K., (5-90 Tomobuchi-cho 1-chome, Miyakojima-ku, Osaka 534-0016, Japan)

A simultaneous and high sensitivity determination method of emedastine and its major metabolites, 5- and 6-hydroxy emedastine, in human plasma was developed. Plasma samples were extracted by Extrelut-1 column. Extracts were separated by reverse phase C18 column under stepwise gradient conditions. An atmospheric pressure chemical ionization interface was used in positive ionization mode. The analytes were detected using a tandem mass spectrometer operated selected reactant monitoring mode. The method was linear over the range 50-5,000 pg/mL (γ=0.999). This method is suitable for therapeutic drug monitoring of unchanged drug and its major metabolites.

Key words: Emedastine, LC/MS/MS, APCI, Human plasma

[ Full Text PDF ]    [ References ]